CAPABILITIES
Continuous, On-Line,
Automatic Monitoring of up
to Eight Sample Lines
Single or Dual Parameters
Multiple Detection Ranges
Multiple Data
Output Capabilities
Up to Eight Concentration
Set Point Alarms
Continual Self Diagnostics
with Alarm
FEATURES
Multiple Wavelength UV
Absorbance Detection System
Integrated Multiple Sample
Line Manifold
Automatic Zero and Clean
Internal Data Logs
No Reagents Required
No Specific Ion Electrodes
to Clean or Replace
Simple to Use
Operator Interface
Sample handling and conditioning
accessories are available for this
and other ChemScan® Analyzers. |
PARAMETERS
Nitrate - Nitrite - Organics
Metals - Color - UV 254
% Transmittance - Monochloramine
See other Models
for additional Parameters

FOR ON-LINE
AND REAL-TIME
WATER AND
WASTEWATER
ANALYSIS
|
APPLICATIONS
Municipal Potable Water
- Surface Water Treatment
- Well Blending
- Ion Exchange Breakthrough
Municipal Wastewater
- Process (with filtration)
- Effluent (without filter)
Cooling/Boiler Water
Industrial Wastewater
Industrial Process Water
- Plating Bath Adjustment
- Leak Detection
BENEFITS
A ChemScan® On-Line Analyzer
can automatically provide
operators with timely process
chemistry measurements,
without the need for frequent
manual sampling or laboratory
analysis. These measurements
can be used to:
Assure Process
Conformance
Control Energy and
Chemical Costs
Improve Process
Performance
|
| TECHNICAL SPECIFICATIONS |
FUNCTIONS AND OUTPUTS
Measurement Principle
Number of Wavelengths
Spectral Range
Calibration Technique
Number of Parameters
Parameter Options
Data Communications
Data Log
Auto Zeroing
Auto Cleaning
Analyzer Pump
Sample Conditioning
Number of Sample Lines
PERFORMANCE SPECIFICATIONS
Reading Interval
Response Time
Range
Accuracy
Precision
Zero Drift
SAMPLE PARAMETERS
Sample Pressure
Sample Flow
Filtration Requirement
Strainer Requirement
Sample Temperature
Sample Turbidity
Sample Suspended Solids
MAINTENANCE
Light Source Replacement
Internal Battery Replacement
Zero/Clean Solution Refill
INSTRUMENT SPECIFICATIONS
Size
Weight
Mounting
Finish Coating
Power
Power Connection
Power Condition
Operator Interface
Sample Cell Material
Sample Connection
Waste Connection
OPERATING ENVIRONMENT
Enclosure Ratings
Ambient Temperature
Relative Humidity
| High Resolution, Ultraviolet Absorbance
256
200 - 450 nm
Pattern Recognition of Spectral Data
1 (Std), 2 max. (Opt)
Primary Absorbance Only (See Table 1)
4 - 2OmA (8 outputs max.), RS-232, other formats optional
1000 Values Time/Date Stamped, 24 Calibration Spectra
YES (Std)
YES (Std)
YES (Std), Zeroing and Cleaning Only
NONE
1 to 8 thru Internal Manifold
1 - 9999 minutes
Immediate after sample flush
Parameter/Site Dependent (See Table 1)
Parameter/Site Dependent
Typ. 2% to 5% of Range
Parameter/Site Dependent
Typ. 2% of Range
Parameter Dependent
Minimized with Auto Zero
10 to 80 psi (Std), 0-10 psi (Opt) w/peristaltic pump
(max. lift 5 ft., max. run 20 ft.)
0.5 to 5 l/min.. (1.5 1 flush/sampling)
NONE (For Samples Meeting Turbidity and Solids Requirements)
Optional ultrafilter available for high solids or turbidity.
Mesh Opening of 2.0 mm Max.
10 - 600C (Std)
0 - 60 NTU (Std)
0 - 150 mg/l TSS
Every 5 years
Every 2 years
As Required (2-4 weeks typ.)
40 x 20 x 10 in.
130 lbs.
Wall (Std) or Stand (Opt)
Baked Enamel on Steel (Std) or Stainless Steel (Opt)
120/240 VAC ±10%, 50-60 Hz, 10 Amps max.
Hard Wired (Std) or Plug (Opt)
Dedicated Branch Circuit Free From: surges/dips > 10%, RF and
Switching Noise
2 x 20 LCD and 4 x 4 Keypad
Polymer (Std) or 316 SS (Opt)
¼" FNPT Fitting
¼" FNPT Fitting (Open Drain Required)
NEMA 4 (Main Enclosure)
NEMA 3R (Optrode Enclosure)
5° - 35°C (Std)
0 - 100% (Non-Condensing)
|
| Table 1 |
| UV-3150 Primary Absorbance Parameters (Standard) |
| Nitrate, Low Range | 0.1 to 10.0 mg/l as NO3 |
| Nitrate, High Range | 10.0 to 100.0 mg/l as NO3 |
| Nitrite, Low Range | 0.1 to 10.0 mg/l as N02 |
| NOx, Low Range | 0.1 to 10.0 mg/l as N |
| NOx, High Range | 10.0 to 100.0 mg/l as N |
| Monochloramine | 0.1 to 10.0 as CI2 |
| Iron III | 0.5 to 20.0 mg/l |
| Molybdenum | 0.5 to 20.0 mg/l |
| Chromium VI | 0.5 to 20.0 mg/l |
| Organics (NOM) | 0.2 to 20.0 mg/l as TOC or COD equivalent |
| % Transmittance | 5.0 to 100.0% |
| UV-254 Absorbance | 0.01 to 2.0 Absorbance Units |
Notes:
1. Technical Specifications are subject to change without prior notice.
2. Parameter range is site dependent and is based on a standard 13mm path length and a potable water matrix.
ChemScan®
Product Installation Drawing #3150
|
Download ASA-3150.pdf
(Adobe Acrobat® format; 182 kbytes)
|
Don't have Acrobat®?

http://www.adobe.com
|
Est. Download Time:
14.4 kbps - 121 seconds
28.8 kbps - 57 seconds
56 kbps - 28 seconds
|
Acrobat is © Adobe Corp.
|
Applications |
Parameters |
Product Selection |
News
Why ChemScan |
Technical Articles |
Openings |
Contact Us
Resources |
F.A.Q. |
Product Literature |
Subscribe
Site Map |
Search Our Site |
Home Page
Applied Spectrometry Associates, Inc.
2325 Parklawn Drive, Suite I · Waukesha, WI 53186
Phone: 262-717-9500 · Fax: 262-717-9530
Toll Free: 800-665-7133
info@chemscan.com
Copyright © Applied Spectrometry Associates, Inc. All Rights Reserved | Privacy Statement
|