Product Literature
New!ChemScan®
UV-2150
Process Analyzers


CAPABILITIES

Continuous, On-Line,
Automatic Monitoring of up
to Four Sample Lines

Single Parameter Analysis

Multiple Detection Ranges

Multiple Data
Output Capabilities

Up to Four Concentration
Set Point Alarms

Continual Self Diagnostics
with Alarm

FEATURES

Multiple Wavelength UV
Absorbance Detection System

Integrated Multiple Sample
Line Manifold

Automatic Zero and Clean

Internal Data Logs

Benign, Inexpensive
Reagents

No Specific Ion Electrodes
to Clean or Replace

Simple to Use
Operator Interface


Sample handling and conditioning
accessories are available for this
and other ChemScan® Analyzers.
PARAMETERS

Ammonia (dissolved)

Phosphate (Total Inorganic, Ortho)

See other Models
for additional Parameters

<b><i>ChemScan<sup>®</sup></i></b> Analyzer
FOR ON-LINE
AND REAL-TIME
WATER AND
WASTEWATER
ANALYSIS
APPLICATIONS

Municipal Potable Water
- Chloramination
- Sequesterant Feed

Municipal Wastewater
- Process (with filtration)
- Effluent (without filter)

Cooling/Boiler Water

Industrial Wastewater

Industrial Process Water


BENEFITS

A ChemScan® On-Line Analyzer
can automatically provide
operators with timely process
chemistry measurements,
without the need for frequent
manual sampling or laboratory
analysis. These measurements
can be used to:

Assure Process
Conformance

Control Energy and
Chemical Costs

Improve Process
Performance


TECHNICAL SPECIFICATIONS
FUNCTIONS AND OUTPUTS
   Measurement Principle
   Number of Wavelengths
   Spectral Range
   Calibration Technique
   Number of Parameters
   Parameter Options
   Data Communications
   Data Log
   Auto Zeroing
   Auto Cleaning
   Analyzer Pump
   Sample Conditioning
   Number of Sample Lines

PERFORMANCE SPECIFICATIONS
   Reading Interval
   Response Time
   Range
   Accuracy

   Precision

   Zero Drift

SAMPLE PARAMETERS
   Sample Pressure

   Sample Flow
   Filtration Requirement
   Strainer Requirement
   Sample Temperature
   Sample Turbidity
   Sample Suspended Solids

MAINTENANCE
   Light Source Replacement
   Internal Battery Replacement
   Zero/Clean Solution Refill
   Reagent Refill

INSTRUMENT SPECIFICATIONS
   Size
   Weight
   Mounting
   Finish Coating
   Power
   Power Connection
   Power Condition
   Operator Interface
   Sample Cell Material
   Sample Connection
   Waste Connection

OPERATING ENVIRONMENT
   Enclosure Ratings

   Ambient Temperature
   Relative Humidity

High Resolution, Ultraviolet Absorbance
256
200 - 450 nm
Pattern Recognition of Spectral Data
One
Secondary Absorbance Only (See Table 1)
4 - 2OmA (4 outputs max.), RS-232, other formats optional
1000 Values Time/Date Stamped, 24 Calibration Spectra
YES (Std)
YES (Std)
YES (Std), Zeroing and Cleaning and Internal Sample Flow Only
YES (Opt), Direct Injection
1 to 4 thru Internal Manifold


3 - 9999 minutes
Analyte Dependent (3 - 5 min. typ)
Analyte/Site Dependent
Analyte/Site Dependent
Typ. 2% to 5% of Range
Analyte/Site Dependent
Less than 0.5% of Range
Analyte/Site Dependent
Minimized with Auto Zero

Pressurized Sample Lines must be regulated to 10 psi maximum,
   with max. lift 5 ft.and max. run 20 ft. to sample location or line connection
0.5 to 5 l/min.. (1.5 1 flush/sampling)
NONE (For Samples Meeting Turbidity and Solids Requirements)
   Optional ultrafilter available for high solids or turbidity.
Mesh Opening of 2.0 mm Max.
10 - 600C (Std)
0 - 60 NTU (Std)
0 - 150 mg/l TSS

Every 5 years
Every 2 years
As Required (2-4 weeks typ.)
As Required (2-4 weeks typ.)


40 x 20 x 10 in.
130 lbs.
Wall (Std) or Stand (Opt)
Baked Enamel on Steel (Std) or Stainless Steel (Opt)
120/240 VAC ±10%, 50-60 Hz, 10 Amps max.
Hard Wired (Std) or Plug (Opt)
Dedicated Branch Circuit Free From: surges/dips > 10%, RF and
  Switching Noise
2 x 20 LCD and 4 x 4 Keypad
Polymer Body with Quartz Windows
¼" FNPT Fitting
¼" FNPT Fitting (Open Drain Required)

NEMA 4 (Main Enclosure)
NEMA 3R (Optrode Enclosure)
5° - 35°C (Std)
0 - 100% (Non-Condensing)


Table 1
UV-2150 Analysis Parameters*
Ammonia
Phosphate (total Inorganic, Ortho)
Chlorine (Free or Total)
Iron (Total)
*Sample Conditioning Reagents Required


Notes:
  1. Technical Specifications are subject to change without prior notice.

ChemScan®
Product Installation
Drawing #2150

Download ASA-2150.pdf
     (Adobe Acrobat® format; 159 kbytes)
Don't have Acrobat®?
Get Acrobat Reader
http://www.adobe.com
Est. Download Time:
14.4 kbps - 106 seconds
28.8 kbps - 57 seconds
56 kbps - 25 seconds
Acrobat is © Adobe Corp.

Applications | Parameters | Product Selection | News
Why ChemScan | Technical Articles | Openings | Contact Us
Resources | F.A.Q. | Product Literature | Subscribe
Site Map | Search Our Site | Home Page

Applied Spectrometry Associates, Inc.
2325 Parklawn Drive, Suite I · Waukesha, WI 53186
Phone: 262-717-9500 · Fax: 262-717-9530
Toll Free: 800-665-7133
info@chemscan.com

Website production & hosting by onyourmark OnYourMark, LLC
Copyright © Applied Spectrometry Associates, Inc. All Rights Reserved | Privacy Statement